Unique Engineering and Instrumentation Solutions

Hard X-Ray DCMs

[photo of 2 APS DCMs]

In the fall of 1996, PSL delivered two unique double crystal monochromators to the Advanced Photon Source at Argonne. The DCMs are the first in the world to operate at ultra-high vacuum and thus can be installed directly on a beamline without a protective window. The DCMs can achieve a vacuum in the 1 × 10-10 Torr range because they are constructed entirely out of aluminum and stainless steel, without any high-torque or ferrofluidic seals and gaskets which would contribute to the contamination of vacuum. Both DCMs are capable of accepting inclined geometry, high heat load optics, and sagittal focusing crystals through provisions of ample space for such optical elements. The basic monochromator is a compact instrument. It has a single rotational stage that does not require a track system to access the crystals yet will fit into the confined back space at APS by having only 300 mm distance between the beam centerline and the back of the instrument. All actuators are hermetically sealed with all services to the actuator routed through a flexible stainless steel corrugated hose. The scan drive uses a stepper motor and glass scale linear encoder. A large flange is located at the front of the chamber through the monochromator can be serviced. Motorized kinematic mounts are used to adjust the position of the entire device.

Features

PARAMETER APS DCM UNI-CAT DCM
Beam offset 35 mm upwards 20 to 40 mm upwards
Energy scan 6 to 24 keV with Si(111) 4 to 40 keV with Si(111)
Range of scan travel 0° to +35° -5° to +45°
Resolution of scan 1 arc sec (5 microradians) 0.5 arc sec (2.5 microradians)
First crystal pitch remote Yes Yes
First Pitch Encoder No Yes
First Pitch range ± 1° ± 3°
First crystal yaw manual Yes
First yaw Range ± 5°
Y1 (normal to crystal) range 0 – 20 mm 0 – 20 mm
Y1 Remotely controllable Yes Yes
Second crystal pitch remote Yes Yes
Second Pitch Encoder No Yes
Second Pitch range ± 1° ± 0.0115°
Second crystal roll remote Yes Yes
Second crystal Roll Encoder No Yes
Second crystal Roll range ± 5° ± 7.5°
Second crystal yaw remote Yes No
Second crystal Yaw Encoder No
Second crystal Yaw range ± 5°
X2 (parallel to pitch axis) No Yes
X2 Encoder Yes
X2 Range 25 mm
Z2 (parallel to crystal surface) Yes Yes
Z2 Encoder Yes Yes
Z2 Range 200 mm 200 mm